This is an open-access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Human Factors, is properly cited. The complete bibliographic information, a link to the original publication on http://humanfactors.jmir.org, as well as this copyright and license information must be included.
Clinical alarm systems safety is a national concern, specifically in intensive care units (ICUs) where alarm rates are known to be the highest. Interventional projects that examined the effect of changing default alarm settings on overall alarm rate and on clinicians’ attitudes and practices toward clinical alarms and alarm fatigue are scarce.
To examine if (1) a change in default alarm settings of the cardiac monitors and (2) in-service nursing education on cardiac monitor use in an ICU would result in reducing alarm rate and in improving nurses’ attitudes and practices toward clinical alarms.
This quality improvement project took place in a 20-bed transplant/cardiac ICU with a total of 39 nurses. We implemented a unit-wide change of default alarm settings involving 17 parameters of the cardiac monitors. All nurses received an in-service education on monitor use. Alarm data were collected from the audit log of the cardiac monitors 10 weeks before and 10 weeks after the change in monitors’ parameters. Nurses’ attitudes and practices toward clinical alarms were measured using the Healthcare Technology Foundation National Clinical Alarms Survey, pre- and postintervention.
Alarm rate was 87.86 alarms/patient day (a total of 64,500 alarms) at the preintervention period compared to 59.18 alarms/patient day (49,319 alarms) postintervention (
Changing default alarm settings and standard in-service education on cardiac monitor use are insufficient to improve alarm systems safety. Alarm management in ICUs is very complex, involving alarm management practices by clinicians, availability of unit policies and procedures, unit layout, complexity and usability of monitoring devices, and adequacy of training on system use. The complexity of the newer monitoring systems requires urgent usability testing and multidimensional interventions to improve alarm systems safety and to attain the Joint Commission National Patient Safety Goal on alarm systems safety in critical care units.
Bedside physiologic monitors are equipped with alarm systems for patient safety and appropriate functionality. Nevertheless, the problematic high volume of false and clinically insignificant nonactionable true positive alarms—up to 99.4%—results in clinicians’ failure to appropriately respond to alarms signaled from monitoring devices [
Alarm safety is a priority in intensive care units (ICUs) where alarm rates are known to be the highest [
After obtaining Institutional Review Board (IRB) approval, this quality improvement project took place in a 20-bed transplant/cardiac ICU located at a university teaching Magnet hospital in the Southwest of the United States. The unit has 39 nurses and an average annual admission of 1500 patients. In April 2014, the unit went through three simultaneous changes, including a move to a new tower and the deployment of new bedside cardiac monitors (Philips IntelliVue MX800) and Cisco phones. The unit is an E shape and has three central nursing stations equipped with cardiac monitors (Philips IntelliVue Information Center iX) with no dedicated monitor watchers. Typical training on new medical devices includes a few hours' demonstration on the appropriate use of devices by the company representative and/or the unit nurse educators. Nurse educators also provide “if needed” support on equipment use. The interventions took place 2.5 months after the move to the new tower and the use of the new cardiac monitors.
Philips IntelliVue MX800 and the Information Center iX cardiac monitors are equipped with complex information systems with tens of main menus and as many submenus, keys, buttons, and icons to facilitate patient data surveillance and management. The monitors are operated using an interactive touch screen, a mouse and a keyboard, or a remote control. The monitors are also capable of complex functions such as lab data integration, drug calculations, guiding care through embedded clinical protocols, issuing reports and strips, presenting trended alarm data, and displaying hundreds of different alarm messages.
The interventions included (1) changing default settings of some parameters on the cardiac monitors and (2) re-educating transplant/cardiac ICU bedside nurses on the appropriate use of the monitors. Default settings were changed based on scientific clinical rational and recent evidence [
1. Limit tightening.
2. Limit increase.
3. Changing the source of alarm detection to enhance alarm reading.
4. Changing the measurement mode in order to capture real conditions from different measurement sources. Measurement modes were changed from “One Source” to “Auto” (eg, System Pulse) and from “Auto” to “Enhanced” (eg, Asystole). “Auto” and “Enhanced” modes allow the monitor to look for an alternate heart rate source, such as the pulse oximeter or the arterial line if it cannot pick up a rhythm from the electrocardiogram (ECG) leads.
5. Alarm delay by increasing the period from alarm detection to announcement.
6. Disabling alarms, for example, Noninvasive Blood Pressure (NBP) Done Tone and Atrial Fibrillation (AFIB). NBP Done Tone is a nonactionable alarm announced automatically by the monitor after measuring the patient blood pressure. The AFIB alarm was disabled because it is also captured by the Irregular Heart Rate alarm. The definitions of alarm events involved in the changes are presented in
Changes were directed toward decreasing the number of false alarms and increasing monitoring safety. For example, although tightening the Premature Ventricular Contractions (PVCs)/minute from 10 bpm to 6 bpm is expected to increase alarm events of this parameter, this tightening was necessary for safety purposes because all other PVC-related alarms, such as Run PVCs, Pair PVCs, Bigeminy, Trigeminy, and Multiform PVCs, were disabled. Similarly, although the limit of ExtremeTachy was increased to decrease the number of false alarms, TachyClamp was tightened for safer monitoring.
7. Volume adjustment including (1) decreasing the volume of yellow alarms with moderate priority, for example, Heart Rate (HR), and (2) increasing the volume of high-priority red alarms, for example, Desaturation. Changes in alarm volume are not expected to directly affect alarm rates, but rather to focus the nurse's attention on actionable high-priority alarms for safety purposes.
All parameters involved in the change are amenable to adjustments by clinicians except for TachyClamp and ExtremeTachy, which are considered hard stops for system safety and can be adjusted only by Philips representatives (see
The nursing unit educators conducted roaming individual in-service sessions. Educational sessions included all nurses in the unit and focused on assessment of monitor parameters, customizing parameters to be patient specific, steps of changing alarm parameters, steps of printing alarm parameters, relearning arrhythmias and changing lead analysis, and troubleshooting common alarming problems (eg, silencing alarms of monitors not connected to patients).
A team of three expert transplant/cardiac ICU nurses and a Philips representative created the list of proposed changes in parameters. This list then went through a review and approval process by all transplant/cardiac ICU physicians, nurse directors, educators, managers, and bedside nurses. The list of approved changes is presented in
After collecting the preintervention surveys, a Philips representative completed a unit-wide change to all monitors based on the approved list on July 1, 2014. This change was also communicated through emails, shift reports, huddles, and meetings to all transplant/cardiac ICU nurses and physicians. The unit in-service education started right after the changes in monitors’ parameters and lasted for approximately two weeks. After that, an invitation to complete the postintervention survey via SurveyMonkey went out to all nurses using individual emails that included the same ID number used in the preintervention survey. Two email reminders were sent to nonrespondents to enhance the response rate.
Alarm events were measured by retrieving the audit log of the cardiac monitors from the database of the central-station monitors for 10 weeks before and 10 weeks after implementing the changes in parameters. The audit log is a chronological record of all alarm events logged by the bedside cardiac monitors.
Nurse characteristics, alarm rate, and attitude toward clinical alarms were described using descriptive statistics.
Changes in default settings of the cardiac monitors at the transplant/cardiac intensive care unit.
Type of change | Parameter | Default setting | Changed to... |
|
|
|
|
|
PVCsa/minute | 10 bpm | 6 bpm |
TachyClampb | 200 bpm | 180 bpm | |
Limit increase | ExtremeTachyb | 20 bpm > Heart Rate High Limit | 40 bpm > Heart Rate High Limit |
|
|
|
|
|
ABPc | Source: Systolic | Source: Systolic and Mean |
NBPd | Source: Systolic and Mean | Source: Systolic | |
|
|
|
|
|
System Pulsee | SpO2f | Auto (from ABP, SpO2, etc) |
Asystole | Standard | Enhanced | |
Alarm delay | SpO2: Averageg | No | Yes: 10 seconds |
|
|
|
|
|
STh Analysisi | On | Off |
|
Run PVCs | On | Off |
|
Pair PVCs | On | Off |
|
Bigeminy PVCs | On | Off |
|
Trigeminy PVCs | On | Off |
|
Multiform PVCs | On | Off |
|
Pause | On | Off |
|
Missed Beat | On | Off |
|
AFIBj | On | Off |
|
NBP Done Tone | On | Off |
Decrease alarm volume | Yellow Alarm Volume | 5 | 3 |
Increase alarm volume | Red Alarm Volume | +0 | +2 |
aPVC: premature ventricular contraction.
bThese alarms are not amenable to change by clinicians. All other alarms can be customized by clinicians based on the patient condition.
cABP: arterial blood pressure.
dNBP: noninvasive blood pressure.
eIf the peripheral capillary oxygen saturation (SpO2) had a poor waveform, the pulse from the pleth would not pick up and would therefore alarm. Changing to Auto allows the monitor to detect a pulse from other sources before alarming.
fSpO2: peripheral capillary oxygen saturation.
gSpO2 will be averaged over 10 seconds to determine a value instead of alarming the second the SpO2 drops below the limit. The nurse can also manually increase this to 20 or 30 seconds.
hST: ST segment in the electrocardiogram.
iThe ST Analysis alarm was disabled but should be turned on for all interventional cardiology cases (eg, require cardiac catheterization) or acute coronary syndrome patients. For these specific patients, the original limit of +/-2.0 mm should be tightened to +/- 1.6 mm as per physicians’ requests.
jAFIB: atrial fibrillation.
Out of the 39 transplant/cardiac ICU nurses who responded to the preintervention survey, 24 (62%) returned completed responses in the postintervention period. General characteristics of the 39 ICU nurses are described elsewhere [
Although all 39 nurses responded to the preintervention survey [
The internal consistency reliability of the 22-item scale that measured attitude toward alarms using Cronbach alpha was high (.72-.75 for the pretest and post-test, respectively).
The positive changes at the postproject period were related to the requirement to document alarm settings, the distinct outputs of medical devices, effective policies to manage alarms, and the ability of the newer systems to solve alarm problems (items 17, 20, 21, and 22). However, in this project, a positive or negative change in attitude on an item was considered clinically meaningful only if reported by at least one-third of nurses (ie, 8 nurses).
Difference in alarm rates between the preproject and postproject periods.
Alarm condition | Preproject period | Postproject period | |||
|
Number of alarms | Total alarm rate/patient day | Number of alarms | Total alarm rate/patient day | |
|
|
|
|
|
|
|
Total | 27,930 | 38.05 | 28,049 | 33.67 |
|
ABPsb (systolic) | 13,776 |
|
14,726 |
|
|
ABPmc (mean) | 13,548 |
|
12,895 |
|
|
ABP disconnect | 606 |
|
428 |
|
Pair PVCsd,e | 8305 | 11.31 | 164 | 0.19 | |
|
|
|
|
|
|
|
Total | 7079 | 9.64 | 8290 | 9.95 |
|
SpO2 | 6741 |
|
7858 |
|
|
SpO2rg (right) | 338 |
|
323 |
|
|
SpO2lh (left) | 0 |
|
109 |
|
Multiform PVCse | 5865 | 7.99 | 19 | 0.02 | |
|
|
|
|
|
|
|
Total | 3686 | 5.02 | 3976 | 4.77 |
|
NBPmj (mean) | 1847 |
|
43 |
|
|
NBPsk (systolic) | 1837 |
|
3933 |
|
|
NBPdl (diastolic) | 2 |
|
0 |
|
PVCs/min | 3233 | 4.40 | 5330 | 6.39 | |
Run PVCs highe | 2155 | 2.94 | 23 | 0.03 | |
STe | 1851 | 2.52 | 2609 | 3.13 | |
|
|
|
|
|
|
|
Total | 1481 | 2.02 | 32 | 0.04 |
|
AFIB | 990 |
|
26 |
|
|
End AFIBn | 491 |
|
6 |
|
Pausee | 1086 | 1.48 | 8 | 0.01 | |
Missed Beate | 873 | 1.19 | 89 | 0.11 | |
Asystole | 323 | 0.44 | 565 | 0.68 | |
|
|
|
|
|
|
|
Total | 292 | 0.39 | 153 | 0.18 |
|
Tachy | 273 |
|
153 |
|
|
Tachy/pp (tachycardia p wave) | 19 |
|
0 |
|
Ventq Bigeminye | 234 | 0.32 | 7 | 0 | |
Vent Trigeminye | 79 | 0.11 | 0 | 0 | |
Pulse | 28 | 0.04 | 5 | 0.01 | |
Total | 64,500 | 87.86 | 49,319 | 59.18 |
aABP: arterial blood pressure.
bABPs: arterial blood pressure systolic.
cABPm: arterial blood pressure mean.
dPVC: premature ventricular contraction.
eThese are the alarms that we disabled.
fSpO2: peripheral capillary oxygen saturation.
gSpO2r: peripheral capillary oxygen saturation right.
hSpO2l: peripheral capillary oxygen saturation left.
iNBP: noninvasive blood pressure.
jNBPm: noninvasive blood pressure mean.
kNBPs: noninvasive blood pressure systolic.
lNBPd: noninvasive blood pressure diastolic.
mAFIB: atrial fibrillation.
nEnd AFIB alarm indicates the end of the AFIB status.
oTachy: tachycardia.
pTachy/p: tachycardia p wave.
qVent: ventricular.
In a previous publication, we reported detailed analysis of the narrative data provided by the 39 transplant/cardiac ICU nurses who responded to the preintervention survey [
In the postintervention survey, 10 out of 24 (42%) nurses provided comments. These comments were matched for the preproject periods and were analyzed. Issues identified were very similar to our previous report [
The respondents’ rankings of the nine statements about the importance of alarm issues specific to cardiac monitors (section 3 in the postintervention survey) is presented in
Number and percentage of nurses who
Item | Statementa | Preproject, |
Postproject, |
% changeb |
1 | Nuisance alarms occur frequently | 24 (100) | 18 (75) | -25.0 |
2 | Nuisance alarms disrupt patient care | 23 (96) | 23 (96) | 0 |
3 | Nuisance alarms reduce trust in alarms and cause caregivers to inappropriately turn alarms off at times other than setup or procedural events | 21 (88) | 22 (92) | 4.8 |
4 | When a number of devices are used with a patient, it can be confusing to determine which device is in an alarm condition | 21 (88) | 19 (79) | -9.5 |
5 | Smart alarms (eg, where multiple parameters, rate of change of parameters, and signal quality are automatically assessed in their entirety) would be effective to use for improving clinical response to important patient alarms | 20 (83) | 17 (71) | -15.0 |
6 | Central alarm management staff responsible for receiving alarm messages and alerting appropriate staff is helpful | 19 (79) | 18 (75) | -5.3 |
7 | Smart alarms (eg, where multiple parameters, rate of change of parameters, and signal quality are automatically assessed in their entirety) would be effective to use for reducing false alarms | 19 (79) | 16 (67) | -15.8 |
8c | Unit layout does interfere with alarm recognition and management | 18 (75) | 18 (75) | 0 |
9 | Alarm integration and communication systems via pagers, cell phones, and other wireless devices are useful for improving alarms management and response | 15 (63) | 17 (71) | 13.3 |
10c | Nearly all alarms are actionable (requiring the nurse to respond and take an action) | 14 (58) | 14 (58) | 0 |
11 | Alarm sounds and/or visual displays of the current monitoring systems and devices should clearly differentiate the priority of alarm | 13 (54) | 14 (58) | 7.7 |
12 | Properly setting alarm parameters and alerts is overly complex in existing devices | 13 (54) | 13 (54) | 0 |
13 | Clinical staff is sensitive to alarms and responds quickly | 13 (54) | 15 (63) | 15.4 |
14c | When a lethal alarm sounds, it is clearly and quickly recognized and immediate action is taken to address the alarm | 12 (50) | 14 (58) | 16.7 |
15 | Environmental background noise has interfered with alarm recognition | 12 (50) | 15 (63) | 25.0 |
16 | Alarm sounds and/or visual displays should be distinct based on the parameter or source (eg, device) | 12 (50) | 16 (67) | 33.3 |
17d | There is a requirement in my unit to document that the alarms are set and are appropriate for each patient | 11 (46) | 18 (75) | 63.6 |
18d | The alarms used on my unit are adequate to alert staff of potential or actual changes in a patient’s condition | 10 (42) | 9 (38) | -10.0 |
19 | There have been frequent instances where alarms could not be heard and were missed | 8 (33) | 8 (33) | 0 |
20d | The medical devices used on my unit all have distinct outputs (ie, sounds, repetition rates, visual displays) that allow users to identify the source of the alarm | 8 (33) | 15 (63) | 87.5 |
21d | Clinical policies and procedures regarding alarm management are effectively used in my unit | 6 (25) | 11 (46) | 83.3 |
22 | Newer monitoring systems (eg, < 3 years old) have solved most of the previous problems we experienced with clinical alarms | 1 (4) | 6 (25) | 500 |
aEdited and used with permission from the Healthcare Technology Foundation (HTF) 2011.
bPercent change = ((y2 - y1) / y1) × 100.
cThese are the new statements that we added to our survey. They do not exist in the original HTF survey.
dThese are the statements where the “floor/area of the hospital” or “institution” in the HTF clinical alarms survey were replaced with “unit” in our survey.
Importance of alarm issues related to the cardiac monitors (n=24).
Item | Statement | Preproject | Postproject |
|
||
|
|
Item response, mean | Mean rankinga | Item response, mean | Mean ranking |
|
1b | Frequent false alarms, which lead to reduced attention or response to alarms when they occur | 2.40 | 1 | 3.40 | 1 | .11 |
2b | Difficulty in understanding the priority of an alarm | 3.00 | 2 | 4.32 | 2 | .07 |
3b | Noise competition from nonclinical alarms and pages | 3.95 | 3 | 4.55 | 3 | .50 |
4c | Lack of available policy on appropriate alarm parameters for individualized patients | 4.40 | 4 | 5.80 | 9 | .08 |
5c | The need to frequently reset alarm settings every time they revert back to default when the monitor is disconnected from the patient | 4.42 | 5 | 5.16 | 5 | .24 |
6d | Difficulty in hearing alarms when they occur, especially from outside patient room | 4.47 | 6 | 5.37 | 7 | .36 |
7d | Difficulty in setting alarms properly because of the complexity of the monitor | 4.84 | 7 | 5.21 | 6 | .70 |
8b | Lack of training on alarm systems | 4.90 | 8 | 4.70 | 4 | .83 |
9d | Difficulty in setting alarms properly because of lack of knowledge on the appropriate limits for my patient condition | 5.42 | 9 | 5.58 | 8 | .75 |
aItem response means were ranked from 1 (most important) to 9 (least important).
bThese statements were adopted from the Healthcare Technology Foundation (HTF) survey.
cThese statements were added to the survey to reflect the cardiac monitors.
dThese statements were modified from the HTF survey. Original statements were as follows: item 6 “Difficulty in hearing alarms when they occur”; items 7 and 9 “Difficulty in setting alarms properly.”
The data showed great variation among nurses in terms of changing alarm parameters (see
The frequency of replacing patients' electrodes also varied. However, only 54% (13/24) of nurses changed them daily during the two project periods (see
In regard to the training needed on the cardiac monitors, the majority of nurses indicated that they did not receive sufficient training on the central and bedside monitors (19/24, 79% and 16/24, 67%, respectively) in the preproject period. Despite the in-service, almost half of the nurses specified their need for more training in the postproject period (see
Percentage of nurses who modify the bedside alarm parameters in the pre- and postproject periods (n=24).
Percentage of nurses who replace patients' electrodes in the pre- and postproject periods (n=24).
Percentage of nurses who received and needed monitor training in the pre- and postproject periods (n=24).
Examining the effect of interventions targeting alarm systems safety on nurses' attitudes toward alarms and alarm fatigue related-practices is critical to evaluate improvements in the safety of these systems. Our unit-wide changes in default alarm settings of cardiac monitors significantly reduced 24% of the total number of the target alarms. However, changing default alarm settings, the subsequent reduction in alarm rate, and the in-service education on alarm management were insufficient to improve nurses’ attitudes toward alarms, alarm fatigue, or maintaining best clinical practices.
Finding alarms for parameters that were turned off supports the fact that bedside nurses customize patient alarms. ABP and SpO2 alarms were among the highest in the preproject and postproject data. The specific types of alarms (eg, Arterial Blood Pressure Mean [ABPm] and Arterial Blood Pressure Systolic [ABPs] alarms; see
The 65% increase in PVCs/min alarms (from 3233 to 5330) is expected because we tightened the parameter. However, the 41% increase in ST alarms (from 1852 to 2609) was unexpected given that we disabled this parameter. Changing this parameter to “On” by nurses is a plausible interpretation for such an increase. The ST parameter includes 12 leads. It would be helpful to analyze if nurses turned on the ST alarm as per the recommended cases by physicians and according to the suggested limits, which leads they adjusted, or if they overused the alarms. Correlating alarm rates and conditions to reliable monitoring conditions is critical and has not yet been investigated. For example, ST monitoring is not recommended in cases when arrhythmias such as atrial flutter and fibrillation are present or if the patient is continuously ventricularly paced. These cases will result in frequent false nonactionable ST alarms.
The unexpected increase in Asystole alarms (from 323 to 565, 75%) can be related to acuity of patients’ conditions and infrequent electrode placement. Another possible explanation from our observation is not adjusting the Pace Mode to “On” in the monitor for patients with temporary pacemakers who keep alarming Asystole. Our results also showed that nurses do not follow the unit protocol (ie, every 24 hours and if needed) when changing leads; enforcement of this policy should take place [
Despite the significant reduction in alarm rate, key issues causing alarm fatigue and reducing trust in alarm systems according to nurses were the high frequency of nuisance alarms, the confusion in locating the alarming device, a unit layout that hinders alarm response, the inadequacy of alarm systems to alert nurses of changes in patients' conditions, the lack of clinical policies and procedures on alarm management, and the complexity of the newer monitoring systems. These multiple issues emphasize the fact that alarm management is very complex in ICUs. On the other hand, and similar to our previous results [
It seems that the complexity of these monitors require interactive, well-designed, and periodic training. Our in-service, though individualized and focused on changing and individualizing alarm parameters and troubleshooting common problems, was insufficient to enhance appropriate monitor use. This is supported by finding that 50% of nurses believed they still needed training on cardiac monitor use and suggests (1) the need for usability testing of cardiac monitors, (2) the use of super-users, and (3) a competency checklist that includes key features for monitor use. Usability studies may reveal the complexity of the monitors, lack of knowledge about some features, or inappropriate use of the monitors. Studies supported the lack of clinicians’ awareness about, and understanding of, the complexity of cardiac monitors [
Cardiac monitors are receiving increased attention in ICUs because of the high number of alarms triggered by these devices compared to other alarm-equipped ICU devices (ie, infusion pumps, dialysis pumps, and mechanical ventilators) [
If we included the alarms from the monitor missed from the unit change and the NBP Done Tone alarms, the actual alarms’ reduction rate would be more than 24%. The inconsistency in applying the same unit of analysis in measuring alarm rates hinders further comparison across alarm safety studies.
The sample of nurses was small. This limited examining the statistical difference in attitudes toward clinical alarms. Although we achieved a significant reduction in alarm rate, we did not correlate that to the acuity of patient conditions preintervention and postintervention. Our description of alarms was limited to the alarms that we targeted for change. The audit log of the cardiac monitors records all types of physiologic alarms and all technical alarms. Analyzing other alarms may provide more insight into the total number of alarms triggered by the cardiac monitors. Our study was limited to alarms from the cardiac monitors and did not include other frequently used alarming devices in ICUs, such as infusion pumps or ventilators. However, cardiac monitors were the devices associated with the highest number of death cases in the US Food and Drug Administration data [
Changing default alarm settings and standard in-service education on cardiac monitor use are insufficient to improve alarm systems safety. Alarm management in ICUs is very complex, involving alarm management practices by clinicians, availability of unit policies and procedures, unit layout, complexity and usability of monitoring devices, and adequacy of training on systems use. The complexity of the newer monitoring systems requires urgent usability testing. Multidimensional interventions are needed to improve alarm systems safety and attain the Joint Commission National Patient Safety Goal on alarm systems safety in critical care units.
Definitions of alarm events for parameters involved in the change.
arterial blood pressure
arterial blood pressure mean
arterial blood pressure systolic
atrial fibrillation
electrocardiogram
heart rate
Healthcare Technology Foundation
intensive care unit
Institutional Review Board
Joint Commission
noninvasive blood pressure
noninvasive blood pressure diastolic
noninvasive blood pressure mean
noninvasive blood pressure systolic
premature ventricular contraction
peripheral capillary oxygen saturation
peripheral capillary oxygen saturation left
peripheral capillary oxygen saturation right
ST segment in the electrocardiogram
tachycardia
tachycardia p wave
ventricular