Assessing the Usability of Six Data Entry Mobile Interfaces for Caregivers: A Randomized Trial

Assessing the Usability of Six Data Entry Mobile Interfaces for Caregivers: A Randomized Trial

Assessing the Usability of Six Data Entry Mobile Interfaces for Caregivers: A Randomized Trial

Original Paper

1Division of Medical Information Sciences, Department of medical imaging and medical information sciences, University Hospitals of Geneva, Geneva, Switzerland

2Division of Anaesthesiology, University Hospitals of Geneva, Geneva, Switzerland

3Division of Clinical Epidemiology, University Hospitals of Geneva, Geneva, Switzerland

4Department of Epidemiology and Preventive Medicine, Monash University, Melbourne, Australia

5Direction of Nursing, University Hospitals of Geneva, Geneva, Switzerland

6School of Medicine, University of Geneva, Geneva, Switzerland

7Faculty of medicine, University of Geneva, Geneva, Switzerland

Corresponding Author:

Frederic Ehrler, PhD

Division of Medical Information Sciences

Department of medical imaging and medical information sciences

University Hospitals of Geneva

Rue Gabrielle-Perret-Gentil 4

Geneva, 1211

Switzerland

Phone: 41 223728697

Fax:41 223728697

Email: frederic.ehrler@hcuge.ch


Background: There is an increased demand in hospitals for tools, such as dedicated mobile device apps, that enable the recording of clinical information in an electronic format at the patient’s bedside. Although the human-machine interface design on mobile devices strongly influences the accuracy and effectiveness of data recording, there is still a lack of evidence as to which interface design offers the best guarantee for ease of use and quality of recording. Therefore, interfaces need to be assessed both for usability and reliability because recording errors can seriously impact the overall level of quality of the data and affect the care provided.

Objective: In this randomized crossover trial, we formally compared 6 handheld device interfaces for both speed of data entry and accuracy of recorded information. Three types of numerical data commonly recorded at the patient’s bedside were used to evaluate the interfaces.

Methods: In total, 150 health care professionals from the University Hospitals of Geneva volunteered to record a series of randomly generated data on each of the 6 interfaces provided on a smartphone. The interfaces were presented in a randomized order as part of fully automated data entry scenarios. During the data entry process, accuracy and effectiveness were automatically recorded by the software.

Results: Various types of errors occurred, which ranged from 0.7% for the most reliable design to 18.5% for the least reliable one. The length of time needed for data recording ranged from 2.81 sec to 14.68 sec, depending on the interface. The numeric keyboard interface delivered the best performance for pulse data entry with a mean time of 3.08 sec (SD 0.06) and an accuracy of 99.3%.

Conclusions: Our study highlights the critical impact the choice of an interface can have on the quality of recorded data. Selecting an interface should be driven less by the needs of specific end-user groups or the necessity to facilitate the developer’s task (eg, by opting for default solutions provided by commercial platforms) than by the level of speed and accuracy an interface can provide for recording information. An important effort must be made to properly validate mobile device interfaces intended for use in the clinical setting. In this regard, our study identified the numeric keyboard, among the proposed designs, as the most accurate interface for entering specific numerical values. This is an important step toward providing clearer guidelines on which interface to choose for the appropriate use of handheld device interfaces in the health care setting.

JMIR Human Factors 2015;2(2):e15

doi:10.2196/humanfactors.4093

Keywords



Electronic data collection and recording in the health care setting is performed increasingly at the patient’s bedside. Data (eg, medical prescriptions, medical summary reports, or daily recordings of body temperature, respiratory or cardiac frequency) can easily be collected on portable computers. Such devices have the advantage of being easy to store, manipulate, and use in emergency departments, outpatient clinics, or other crowded areas. Among the portable devices on offer, tablets and smartphones are becoming increasingly popular due to their handiness and resemblance to traditional paper-and-pencil data collection interfaces [1-3]. They also offer the advantage of providing apps designed especially for handheld devices, such as drug dosage calculators, electronic pharmacopeias, textbooks, or medical literature databases [1,4-7].

In the health care setting, many usability problems contribute to medical errors [8], of which those related to data entry are a major source. The quality of recorded information is of utmost importance because the life of a patient can easily be put at risk by the improper recording of a drug dosage or the incorrect labeling of a physiological or biological value [9-11]. Accuracy in the process of data recording can be significantly influenced by the design of an interface or by factors related to the type of data to be recorded, such as a number’s length, type, magnitude or frequency, and even font appearance [12]. This has already been demonstrated for specific entry devices, such as infusion pumps [12-14], and in the context of medical prescriptions [8]. The limited size and tactile interaction paradigm of handheld devices compared to traditional laptops has further emphasized the risk of increased errors in data entry [3].

The influence that specific characteristics of handheld devices can have on human-machine interactions has already been studied [15-27]. Earlier works investigated the impact of limited display size on users’ performances for tasks such as browsing, information retrieval, readability, or target selection [15-19]. These studies showed that the size of a mobile screen has no major impact on the user’s comprehension of the information displayed. However, a correlation was found between the ease of reading and the size of the screen. In this context, Duchnicky and Kolers [20] found that it takes users up to 25% longer to read a given text on a small-width display than on a regular desktop screen width. This was confirmed by another study by Resiel and Shneiderman [21] who reported that reading on a 22-line display compared with a 60-line display resulted in a 15% decrease in the speed of reading. The same is true for information retrieval; studies have indicated that information retrieval tasks are harder to perform on small screen devices [22]. Moreover, users are more likely to perform incorrect choices when selecting from possible links and waste more time carrying out additional scrolling type activities [23,24]. Screen size also influences the quality of information input. Most studies assessing users’ performance have confirmed that data input accuracy can be impacted by keyboard size or character setting, but also by other factors, such as a user’s finger size [25-27]. Generally, a familiar disposition of the display and large keyboards improve user performance.

All these tend to demonstrate that it is crucial to take into account screen size specificity in the design of mobile device interfaces. Therefore, we assessed 6 tactile interfaces representing many of the common interfaces used on tablet PC and smartphones as interfaces built based on user requirements.


In order to identify the most suitable interface for the effective and reliable recording of numerical data on handheld devices, we performed a crossover randomized trial assessing 6 handheld interfaces designed according to 6 paradigms ranging from commercially available solutions to experimental designs. Each participant had to record several vital numerical signs on each interface. The interfaces were provided automatically in a random order.

Participants

A sample most representative of health care professionals likely to collect, record, or use clinical data as part of their daily activity was recruited within the University Hospitals of Geneva, Switzerland. Recruitment was carried out following approval of the project by the “Commission cantonale d'éthique de la recherche,” the University hospitals’ human research and ethics committee. Previously published research on similar research questions were able to demonstrate significant differences between data entry interfaces with 30 participants involved [13]. However, because we were able to recruit among a large number of caregivers and had no prerequisites as to participants’ characteristics, we relied on a convenience sampling approach. In total, 150 hospital workers selected across 5 categories of professionals were recruited: nurses, assistant nurses, midwives, physicians, and administrative staff. Participation was on a voluntary basis and there was no exclusion criterion.

Instruments

To test the 6 interfaces, we used a common commercially available smartphone, the Samsung Galaxy Note, which has a 5.3-inch screen with a resolution of 1280×800 pixels. This type of smartphone is representative of contemporary smartphones. It is also characterized by a high level of flexibility, which reduces to the minimum the design constraints set by human-machine interfaces. The 6 interfaces that were tested were designed to represent the most complete range of options man-machine computer interfaces can offer (Figure 1). Four of the interfaces were chosen among existing interface design frameworks. The numeric keypad interface was chosen because it is recognized as being a being very effective at reducing the number of entry errors [13]. The stepper and wheel interfaces were chosen because they are commonly available on 2 major commercial smartphone operating systems, Android and iOS. Finally, the character recognition interface was chosen because it was among the first interfaces proposed for touchscreen devices, such as the PalmPilot. The last two options were developed by two distinct expert committees based on end-user requirements for recording interfaces. The first committee included caregivers who designed a dynamic chart (named “column”) aimed at facilitating the identification of vital sign trends. The second committee included computer scientists at the hospital who developed a fast data entry system (the “circle”) built on an interaction principle similar to the SwiftKey keyboard. The interfaces were implemented by the research team and installed on all devices provided to participants. Experimental designs were previously extensively tested by volunteers to ensure usability of the recording interfaces.

Figure 1. Description of the six data entry interfaces.
View this figure

Study Procedure and Data Collection

The trial was designed to minimize interaction biases. Based on a set of possible problems identified in a pilot phase, standardized procedures were defined so that any problem (eg, a system crash) would by default always be handled in a similar way. Each session of the trial took place in a controlled environment where participants could perform the test without any interruption or other disturbance. Before beginning the trial, each participant was shortly briefed on the study purpose and overall procedure. Participants were also informed that there was no time constraint for entering the displayed numbers. Each participant was then provided with a smartphone. Because the study procedure and instructions were available on each device and the entire study process was regulated by a computer program, there was no need for further interaction with the research team. Although the experiment took place in a controlled environment, participants were asked to use only one hand to hold the device to simulate the real-life bedside recording procedure as closely as possible (see Multimedia Appendix 1 for the CONSORT checklist filled for this study).

Figure 2. Data recording process.
View this figure

Participants were first asked to read the instructions of the experiment that were displayed on the smartphone. A short questionnaire then popped up on the screen asking for details on participants’ demographics and computer literacy. Following these steps, the actual trial phase began. Each participant was requested to enter data on all 6 interfaces. The sequence in which interfaces came up was automatically computed and randomly defined for each participant. The procedure was the same for each interface: participants were first able to practice the procedure before entering the data to be recorded. For each interface design, participants were asked to enter 3 types of physiological measures (body temperature, respiratory rate, pulse rate) that were displayed at the top of the screen. We chose these 3 vital signs from among the most frequently collected signs during patient follow-up that could be described using a single numerical value. For this reason, measurements such as blood pressure (composed of 2 values) were discarded as were other types of data, such as dates or time, which we considered would require other types of dedicated entry interfaces. When entering a number, participants could correct it as many times as they wanted before validating it. The computer program randomly generated the type of physiological data to be recorded and a value to be entered. During the training period, participants had to record 2 physiological values with the interface displayed on the screen. Users could not skip this step and had to continue the selection process until they succeeded. This is the only time users were allowed to request external assistance if they did not understand how to use the interface. Once the 2 physiological values were successfully recorded, the testing process started and participants were asked to record 3 random values for each possible interface-sign combination. Data for pulse, respiratory rate, and body temperature were given a predefined range of recordable values (Table 1).

Table 1. Range of recordable values.
SignRangeDecimalNumber of possible values
Pulse30-170No140
Body temperature36-41Yes50
Respiratory rate3-20No17

The task was repeated until 3 values were recorded for each of the 3 physiological signs in each of the 6 interfaces. Altogether, this totaled 54 data entries for each participant.

Measured Variables

All variables needed for assessing the performance of each participant were recorded automatically on the smartphone. Among the variables that were measured were the number of actions performed by the participant, the number of corrections made, and the time until data entry validation. Data could be exported in a comma separated value file. Accuracy was measured by comparing the values participants were requested to enter as they appeared on the top of the smartphone’s screen with the ones actually entered by each participant. Success or failure outcomes were then summed and reported as a proportion of the maximum possible score for each category of data entry. Effectiveness was assessed by measuring the time used to record each of the values.

Analysis

Descriptive summary statistics of continuous variables included means (SD), or medians with ranges, depending on their distribution. Continuous variables were compared using the paired Student t test or the Wilcoxon signed rank test if not normally distributed. For categorical variables, we used frequencies and proportions. Participants’ satisfaction was assessed using the Friedman 2-way ANOVA test. After adjusting for age and familiarity of use of computers or smartphones, possible associations between the data entry time for each type of variable and the type of interface used or the sequence of data entry were examined using multilevel linear models. The same was done to evaluate the number of errors and corrections associated with each interface, this time using multilevel mixed logistic models. To obtain a normal distribution of the dependent variable, we used the log of speed of data entry. Interfaces were nested within the type of data entry, which were themselves nested within the coder.

A P value <.05 was considered statistically significant. We performed all analyses using the statistical package Stata version 12.


Only 4 participants failed to complete the test due to technical problems or professional emergency causing them to leave the study before its completion. These incomplete records were removed from the study analysis. Demographics for the remaining 146 participants could be summarized as follows: mean age of participants was 43.6 years (SD 10.2) with two-thirds (64.4%, 94/146) younger than age 46 years. Most of participants enrolled in the study (92.5%, 135/146) were right-handed and nearly all (95.9%, 140/146) possessed a personal computer, although only 61.0% (89/146) already owned a smartphone. Most (75.3%, 110/146) were caregivers; the remaining 24.7% (36/146) were recruited among administrative personnel or computer scientists classified as “other participants.” Study participant characteristics are detailed in Table 2.

The time needed to enter values with the different interfaces differed depending on the type of vital sign recorded (Table 3). In most cases, it took less time to enter pulse than body temperature values, and even less time to enter respiratory rates (except for the wheel mode). The results provided in Table 3 show that the speed of data entry was also influenced by the type of interface used. Differences were statistically significant (P<.001). The fastest interface for data entry was the numeric keyboard, with a mean entry speed ranging from 2.81 (SD 0.06) to 4.34 (SD 0.08) seconds depending on the type of variable recorded. This was significantly quicker than the stepper (mean 4.31, SD 0.11 seconds to mean 7.23, SD 0.36 seconds), the wheeler (mean 5.13, SD 0.14 seconds to mean 8.35, SD 0.21 seconds), and the circle (mean 4.45, SD 0.01 seconds to mean 9.38, SD 0.28 seconds) models. Finally, the less efficient interfaces were the column (mean 5.25, SD 0.16 seconds to mean 10.49, SD 0.45 seconds) and character recognition (mean 6.32, SD 0.24 seconds to mean 14.68, SD 0.64 seconds) models.

Table 4 shows data entry accuracy for the different interfaces. The numeric keyboard was the most accurate interface (96.3%-99.3% accuracy). It was followed by the stepper (97.9%-98.4% accuracy) and the wheel (95.4%-98.6% accuracy). Other interfaces were associated with more data recording errors. The other interfaces yielded levels of accuracy ranging from 93.2% to 94.3% for the column model, 88.6% and 96.1% for the circle, and 81.5% to 86.8% for the character recognition interface.

Among the tested interfaces, the numeric keyboard achieved the highest level of accuracy for recording pulse data. When looking at the mean scores of all 3 measures, the overall performance of the numeric keyboard was comparable to that of the stepper and wheel models. With regard to speed of data entry, the numeric keyboard scored the highest, with results at least 1.5 times faster than all other interfaces regardless of the type of variable recorded. Interfaces built based on a participative design, such as the circle or column, were associated with additional errors. Character recognition was found to be the most inaccurate and slowest interface for clinical data recording (Table 5).


These study results can be explained by several factors. First, the selection mechanism used to enter values into the system was the most influential factor for determining speed of data entry. With a deterministic selection mechanism, such as the keyboard, the actions performed by users are transformed unambiguously into the associated outcome. This differs from a nondeterministic selection mechanism, such as character recognition, where there is no guarantee that the action of the user is transformed into the desired outcome. The interfaces with such nondeterministic selection mechanisms require some level of expertise and training to perform accurate actions that can be transcribed by the system into the desired outcome. The number “1” could, for instance, be at misinterpreted as a “7” and therefore require additional time to be corrected. The second parameter influencing the speed of data entry is the number of actions required to record it. In this regard, the stepper is not as fast as the keyboard because it requires more actions to enter a given numerical value.

An analysis of the error rate enabled us to classify the interfaces into 2 main categories: those yielding a level of accuracy greater than 95% (numeric keyboard, stepper, and wheel) and those yielding levels less than 95% (character recognition, circle, and columns). Interfaces offering no immediate feedback and those where it was difficult to modify recordings (character recognition, wheel) were associated with a significantly increased risk of recording errors.

Table 2. Participant characteristics (N=146).
Participant characteristicsn (%)
Age (years)

22-3442 (28.8)

35-4552 (35.6)

≥4652 (35.6)
Gender

Female96 (65.7)

Male50 (34.3)
Hand preference

Left-handed11 (7.5)

Right-handed135 (92.5)
Profession

Assistant nurse4 (2.7)

Nurse76 (52.2)

Administrative19 (13.0)

Midwife11 (7.5)

Physician19 (13.0)

Other17 (11.6)
Has a computer

No6 (4.1)

Yes140 (95.9)
Has a smartphone

No57 (39.0)

Yes89 (61.0)
Frequency of vital sign recording

Never45 (30.8)

Monthly12 (8.2)

Every day89 (61.0)
Table 3. Mean time to enter 3 variables for each interface-sign combination.
Data typeInterface entry time (sec), mean (SD)P

Character recognitionCircleColumnsNumeric keyboardStepperWheel
Pulse10.50 (0.44)6.27 (0.14)10.06 (0.34)3.08 (0.06)6.53 (0.36)8.35 (0.21)<.001
Body temperature14.68 (0.64)9.38 (0.28)10.49 (0.45)4.34 (0.08)7.23 (0·18)7.22 (0.19)<.001
Respiratory rate6.32 (0.24)4.45 (0.01)5.25 (0.16)2.81 (0.06)4.31 (0.11)5.13 (0.14)<.001
Table 4. Correct data entries for the 3 variables on each of the 6 interfaces.
Data typeCorrect data entries, n (%)P

Character recognitionCircleColumnsNumeric keyboardStepperWheel
Pulse372 (84.9)421 (96.1)413 (94.3)435 (99.3)431 (98.4)432 (98.6)<.001
Body temperature357 (81.5)388 (88.6)412 (94.1)425 (97.0)431 (98.4)427 (97.5)<.001
Respiratory rate380 (86.8)413 (94.3)408 (93.2)422 (96.3)429 (97.9)418 (95.4)<.001
Table 5. Data entry speed and accuracy for different interface characteristics.
Interface characteristicsMean time (sec)Mean accuracy, %Minimum number of actions measuredSelectionDesign source
Numeric keyboard3.4197.51 per digitDeterministicLiterature
Stepper6.0298.2Sum of the digitsDeterministicAndroid OS
Circle6.7931 per digitDeterministicComputer scientists
Wheeler6.997.21 per numberNondeterministiciOS
Columns8·693.91 per numberNondeterministicCaregivers
Character recognition10.584.41 per digitNondeterministicPalm OS

Our results are in line with other research findings [13,28,29]. A study comparing 5 number entry interfaces for an infusion pump [13] showed lower error rates and speed of data entry when entering values on a number pad rather than a stepper. In a study evaluating data recording accuracy of a keyboard for electronic health recording systems compared with handwritten medical paper records, authors found that 25.6% of vital signs had one or more errors when documented on paper medical records compared to 14.9% errors when documented in an electronic format [28]. This confirms the improved accuracy of a keyboard compared to handwriting, as used in character recognition interfaces. Another study by Wager et al [29] compared error rates for 4 different types of entry devices (paper, computer on wheel, tablet, and direct feed from the monitoring system). Although no details were provided about the interfaces used, the error rate associated with a keyboard on a tablet was 5.6%, on average, similar to our study findings.

There are several limitations to our study. The first is the lack of exact knowledge about the prior level of familiarity users had for each of the 6 interfaces. However, one can reasonably assume that the novel interfaces implemented for the study were unknown to all users, whereas numeric keyboard and character recognition interfaces were probably familiar to computer and mobile phone users. Our study sample was mostly composed of the latter (see Table 2). In order to minimize this bias, we included a training session for each interface in our protocol. It is unclear whether this was sufficient to ensure a similar level of mastery for all categories of interfaces, but this limited the risk of variability due to insufficient training rather than to a true difference between interfaces [30,31].

A second limitation is the unknown influence of the original parameters set into the system on study outcomes [32,33]. Indeed, there are many factors that can influence the data entry process. For instance, the size of the keyboard buttons is known to influence a user’s performance during data entry [27,34]. On infusion pumps, such factors can generate up to 28 different implementations of the data entry interface [13]. In their study, Cauchi et al [35] confirmed that these parameters not only influence data entry performance, but also the severity/degree of recording errors. Likewise, the character recognition algorithm can facilitate or complicate user inputs [36]. Therefore, if our results show clear and statistically sound differences in interface usability and accuracy, other studies analyzing the same interfaces but using different settings may show different results.

Third, despite explicitly instructing participants to act as they would normally during the trial session, while at the same time forbidding observers to intervene after the testing phase, it cannot be excluded that external observers may have had an influence on participants’ behavior. For example, participants may have improved or modified an aspect of their behavior in response to the context in which they were acting rather than in response to the type of device used. This is known as the “Hawthorne effect” [37].

Fourth, the experimental setting cannot be considered as an ecologically valid one because the experiment took place in a controlled environment. Although we attempted to minimize this limitation by asking users to carry the device in only one hand, the fact that the numbers to enter were provided directly on the smartphone did not reflect real-life situations, where numbers are more likely to be read from another source. We chose this method of data delivery to exclude errors that might otherwise arise when transcribing data from one place to another [38].

Fifth, the purely quantitative approach adopted in our study cannot provide explanations as to why variations exist across different interfaces and what the implications are for the design and evaluation of mobile data entry tools. To clearly answer these questions, a more qualitative methodology needs to be adopted, such as the one used by Kushniruk et al [8] in their analysis of the relationship between usability problems and prescription errors in handheld applications.

Finally, it is also worth pointing out that evaluating failure based on a binary measure does not make it possible to evaluate the type and potential severity of an error. Indeed, studies such as those carried out by Wiseman et al [39] or Oladimeji et al [13,40] reported up to 7 classes of errors. Due to this limitation, our results do not highlight the fact that interfaces that enable each digit to be chosen independently are much more likely to create errors of higher magnitude and usually critical in a clinical context [12,41,42].

Although there is a marked increase in use of handheld devices in the health care environment, not all interfaces are adapted to specific constraints, such as smaller, interactive tactile screens. The possibility of recording clinical information on mobile devices opens the door to many questions. Interfaces set on these devices are usually complex to use. In his study, Howard [43] showed that this complexity could be explained by 4 reasons: intricacy, equivalence, omniscience, and commitment. To account for these different aspects, the design of a new interface should evaluate each parameter separately for effectiveness and accuracy before combining parameters into a more complex design. Moreover, the identification of the most appropriate handheld user interface to record numerical data in an electronic format is only one of the many aspects that need to be investigated. Further aspects, such as the manipulation of text data or the recording of graphical data, need to be analyzed also.

Recording information in real time on handheld devices at the patient’s bedside is increasingly becoming the standard of care in many health care settings. These devices offer improved portability and flexibility compared to desktop computers. The ease and accuracy with which data can be recorded in such settings will determine the choice of the most appropriate human-machine interface. Although a lot of work has been done on physical keyboards by engineers and ergonomists to improve the reliability and efficiency of data recording, a lot of work still has to be done for handheld devices. Our study shows that among the interface designs we selected, the keyboard reached the highest level of speed and accuracy when recording pulse data. The stepper and wheeler interfaces demonstrated similar accuracy, whereas the numeric keyboard remained by far the quickest interface for tactile interaction. Whether this conclusion will remain valid when using other interface parameters remains to be tested; however, the need for more in-depth evaluation of novel interfaces has clearly been demonstrated.

Although some interface designs may, at first sight, appear promising, only formal and rigorous assessments in randomized trials will enable the identification of the most accurate and usable interfaces for data recording in the clinical setting. As the new generation of handheld devices progressively replaces traditional computers, future developments to find the most appropriate human-machine interface should not only be based on designer or user committee inputs, but also on advice from other fields, such as experts in ergonomics.

Acknowledgments

This work has been funded by the University Hospitals of Geneva.

Conflicts of Interest

The authors are owners or developers of the software used in this study.

Multimedia Appendix 1

CONSORT checklist.

PDF File (Adobe PDF File), 897KB

  1. Lu Y, Xiao Y, Sears A, Jacko JA. A review and a framework of handheld computer adoption in healthcare. Int J Med Inform 2005 Jun;74(5):409-422. [CrossRef] [Medline]
  2. Fischer S, Stewart TE, Mehta S, Wax R, Lapinsky SE. Handheld computing in medicine. J Am Med Inform Assoc 2003;10(2):139-149 [FREE Full text] [Medline]
  3. Haller G, Haller DM, Courvoisier DS, Lovis C. Handheld vs. laptop computers for electronic data collection in clinical research: a crossover randomized trial. J Am Med Inform Assoc 2009;16(5):651-659 [FREE Full text] [CrossRef] [Medline]
  4. Mosa AS, Yoo I, Sheets L. A systematic review of healthcare applications for smartphones. BMC Med Inform Decis Mak 2012;12:67 [FREE Full text] [CrossRef] [Medline]
  5. Prgomet M, Georgiou A, Westbrook JI. The impact of mobile handheld technology on hospital physicians' work practices and patient care: a systematic review. J Am Med Inform Assoc 2009;16(6):792-801 [FREE Full text] [CrossRef] [Medline]
  6. Wu RC, Straus SE. Evidence for handheld electronic medical records in improving care: a systematic review. BMC Med Inform Decis Mak 2006;6:26 [FREE Full text] [CrossRef] [Medline]
  7. Kuziemsky CE, Laul F, Leung RC. A review on diffusion of personal digital assistants in healthcare. J Med Syst 2005 Aug;29(4):335-342. [Medline]
  8. Kushniruk A, Triola M, Borycki E, Stein B, Kannry JL. Technology induced error and usability: the relationship between usability problems and prescription errors when using a handheld application. Int J Med Inform 2005 Aug;74(7-8):519-526. [CrossRef] [Medline]
  9. Masci P, Zhang Y, Jones P, Thimbleby H. A Generic User Interface Architecture for Analyzing Use Hazards in Infusion Pump Software. In: MCPS 2014. 2014 Apr 10 Presented at: 5th Workshop on Medical Cyber-Physical Systems; Apr 14, 2014; Berlin, Germany p. 1-14. [CrossRef]
  10. Fairbanks RJ, Caplan SH, Bishop PA, Marks AM, Shah MN. Usability study of two common defibrillators reveals hazards. Ann Emerg Med 2007 Oct;50(4):424-432. [CrossRef] [Medline]
  11. Garmer K, Liljegren E, Osvalder A, Dahlman S. Application of usability testing to the development of medical equipment. Usability testing of a frequently used infusion pump and a new user interface for an infusion pump developed with a Human Factors approach. International Journal of Industrial Ergonomics 2002 Mar;29(3):145-159. [CrossRef]
  12. Tu H, Oladimeji P, Li Y, Thimbleby H, Vincent C. The dffects of number-related factors on entry performance. In: BCS-HCI '14 Proceedings. 2014 Sep 09 Presented at: 28th International BCS Human Computer Interaction Conference on HCI; September 9-12, 2014; Southport, UK p. 8221-8251.
  13. Oladimeji P, Thimbleby H, Cox A. A performance review of number entry interfaces. In: Lecture Notes in Computer Science. 2013 Sep 2 Presented at: Human-Computer Interaction–INTERACT 2013; September 2-6, 2013; Cape Town, South Africa p. 365-382. [CrossRef]
  14. Li K, Ding S, Dong Z, Qin L. MediCHI: safer interaction in medical devices. 2013 Apr 27 Presented at: CHI’13; Apr 27-May 2, 2013; Paris, France p. 3267-3270. [CrossRef]
  15. Kwon S, Kim C, Kim S, Han S. Two-mode target selection: considering target layouts in small touch screen devices. International Journal of Industrial Ergonomics 2010 Nov;40(6):733-745. [CrossRef]
  16. Zimmerman D, Yohon T. Small-screen interface design: Where are we? Where do we go? 2009 Sep 19 Presented at: 2009 IEEE Int Prof Commun Conf IEEE; Jul 19-22, 2009; Waikiki, HI p. 1-5. [CrossRef]
  17. Alghamdi E, Yunus F, Househ M. The impact of mobile phone screen size on user comprehension of health information. Stud Health Technol Inform 2013;190:154-156. [Medline]
  18. Ziefle M. Information presentation in small screen devices: the trade-off between visual density and menu foresight. Appl Ergon 2010 Oct;41(6):719-730. [CrossRef] [Medline]
  19. Dillon A, Richardson J, Mcknight C. The effects of display size and text splitting on reading lengthy text from screen. Behaviour & Information Technology 1990 May;9(3):215-227. [CrossRef]
  20. Duchnicky RL, Kolers PA. Readability of text scrolled on visual display terminals as a function of window size. Hum Factors 1983 Dec;25(6):683-692. [Medline]
  21. Reisel J, Shneiderman B. Is bigger better?: The effects of display size on program reading. 1987 Aug 10 Presented at: Social, Ergonomic and Stress Aspects of Work with Computers; Aug 10-14, 1987; Honolulu, Hawaii p. 113-122.
  22. Kim L, Albers MJ. Web design issues when searching for information in a small screen display. 2001 Presented at: SIGDOC '01 19th annual international conference on Computer documentation; Oct 21-24, 2001; New York p. 193-200.
  23. Jones S, Jones M, Marsden G, Patel D, Cockburn A. An evaluation of integrated zooming and scrolling on small screens. International Journal of Human-Computer Studies 2005 Sep;63(3):271-303. [CrossRef]
  24. Jones M, Marsden G, Mohd-Nasir N, Boone K, Buchanan G. Improving Web interaction on small displays. Computer Networks 1999 May;31(11-16):1129-1137. [CrossRef]
  25. Coleman MF, Loring BA, Wiklund ME. User performance on typing tasks involving reduced-size, touch screen keyboards. 1991 Presented at: Vehicle Navigation and Information Systems Conference; Oct 20-23, 1991; New York p. 543-549. [CrossRef]
  26. Siek K, Rogers Y, Connelly K. Fat finger worries: how older and younger users physically interact with PDAs. In: Proceedings of the 2005 IFIP TC13 international conference on Human-Computer Interaction. 2005 Presented at: INTERACT'05 IFIP TC13 international conference on Human-Computer Interaction; Sep 12-16, 2005; Rome p. 267-280. [CrossRef]
  27. Sears A, Revis D, Swatski J, Crittenden R, Shneiderman B. Investigating touchscreen typing: the effect of keyboard size on typing speed. Behaviour & Information Technology 1993 Jan;12(1):17-22. [CrossRef]
  28. Gearing P, Olney CM, Davis K, Lozano D, Smith LB, Friedman B. Enhancing patient safety through electronic medical record documentation of vital signs. J Healthc Inf Manag 2006 Jan;20(4):40-45. [Medline]
  29. Wager KA, Schaffner MJ, Foulois B, Swanson Kazley KA, Parker C, Walo H. Comparison of the quality and timeliness of vital signs data using three different data-entry devices. Comput Inform Nurs 2010;28(4):205-212. [CrossRef] [Medline]
  30. Anderson AM, Mirka GA, Joines SM, Kaber DB. Analysis of alternative keyboards using learning curves. Hum Factors 2009 Feb;51(1):35-45. [Medline]
  31. Pirhonen A. What do learning curves tell us about learnability. In: Proceedings of the Human Factors Conference HF2002. 2002 Nov 25 Presented at: Human Factors Conference HF2002; Nov 25-27, 2002; Melbourne.
  32. Harrison R, Flood D, Duce D. Usability of mobile applications: literature review and rationale for a new usability model. J Interact Sci 2013;1(1):1-16. [CrossRef]
  33. Hornbæk K. Current practice in measuring usability: Challenges to usability studies and research. International Journal of Human-Computer Studies 2006 Feb;64(2):79-102. [CrossRef]
  34. Sears A, Zha Y. Data entry for mobile devices using soft keyboards: understanding the effects of keyboard size and user tasks. International Journal of Human-Computer Interaction 2003 Oct;16(2):163-184. [CrossRef]
  35. Cauchi A, Gimblett A, Thimbleby H, Curzon P, Masci P, Cauchi A. Safer "5-key" number entry user interfaces using differential formal analysis. 2012 Sep 12 Presented at: 26th Annual BCS Interaction Specialist Group Conference on People and Computers; Sep 12-14, 2012; Birmingham, UK p. 29-38.
  36. Lecun Y, Jackel L, Bottou L, Cortes C, Denker J, Drucker H, et al. Learning algorithms for classification: a comparison on handwritten digit recognition. 1995 Presented at: Neural Networks: The Statistical Mechanics Perspective; Jan 1995; New York p. 261-276.
  37. McCarney R, Warner J, Iliffe S, van Haselen R, Griffin M, Fisher P. The Hawthorne Effect: a randomised, controlled trial. BMC Med Res Methodol 2007;7(30) [FREE Full text] [CrossRef] [Medline]
  38. Li Y, Oladimeji P, Thimbleby H. Exploring the effect of pre-operational priming intervention on number entry errors. 2015 Apr 18 Presented at: 33rd Annual ACM Conference on Human Factors in Computing Systems; Apr 18-23, 2015; Seoul, Korea p. 1335-1344. [CrossRef]
  39. Wiseman S, Cairns P, Cox A. A taxonomy of number entry errors. 2011 Presented at: 25th BCS Conference on Human-Computer Interaction; Jul 4-8, 2011; Swinton, UK p. 187-196.
  40. Oladimeji P, Thimbleby H, Cox A. Number entry interfaces and their effects on error detection. In: Proceedings of INTERACT 2011 Human-Computer Interaction Conference. Heidelberg: Springer; 2011 Sep 5 Presented at: INTERACT 2011: 13th IFIP TC 13 International Conference; Sep 5-9, 2011; Lisbon, Portugal.
  41. Thimbleby H, Cairns P. Reducing number entry errors: solving a widespread, serious problem. J R Soc Interface 2010 Oct 6;7(51):1429-1439 [FREE Full text] [CrossRef] [Medline]
  42. Oladimeji P. Towards safer number entry in interactive medical systems. 2012 Jun 25 Presented at: 4th ACM SIGCHI symposium on Engineering interactive computing systems; Jun 25–28, 2012; Copenhagen, Denmark p. 329-332. [CrossRef]
  43. Howard S. Trade-off decision making in user interface design. Behaviour & Information Technology 1997 Jan;16(2):98-109. [CrossRef]

Edited by G Eysenbach; submitted 02.12.14; peer-reviewed by S Bradley, C Kuziemsky, P Curzon; comments to author 17.04.15; revised version received 06.07.15; accepted 04.11.15; published 15.12.15

Copyright

©Frederic Ehrler, Guy Haller, Evelyne Sarrey, Magali Walesa, Rolf Wipfli, Christian Lovis. Originally published in JMIR Human Factors (http://humanfactors.jmir.org), 15.12.2015.

This is an open-access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Human Factors, is properly cited. The complete bibliographic information, a link to the original publication on http://humanfactors.jmir.org, as well as this copyright and license information must be included.